Back
Browse Products By:
APPLICATIONS
Back
Browse Products By Application
View All Applications

XRB100N100 Monoblock® Series

  • Integrated HV Supply, Filament Supply, X-Ray Tube, Beam Port and Control Electronics
  • Compact & Lightweight
  • Universal input, Power Factor Corrected with Internal EMI Filter
  • Can be Mounted in Any Physical Orientation
  • Analog Control Interface and Standard RS-232 Digital Interface
Get a Quote

Spellman’s XRB100N100 Monoblock® X-Ray source is designed for OEM applications powering its internal X-Ray tube up to 100kV at 100W. Features like universal input, small package size and a standard analog and RS-232 digital interface simplify integrating this Monoblock® into your X-Ray system. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability performance.

Typical applications:

  • X-Ray Scanning: Food Inspection
  • Fill Level Confirmation
  • Security Applications
  • Specifications
  • Downloads

Specifications

(Ref. 128072-001 REV. L)

  • X-Ray Characteristics:
  • Tube Type:
    Stationary anode, tungsten target
  • Focal Spot:
    0.5mm (IEC 336)
  • Beam Filter:
  • Lexan:
    3.2mm
  • Oil:
    10mm ±0.1mm
  • Glass:
    1.8mm max
  • Beam Geometry:
    Symmetrical fan 74° x 10° ±1°
  • Input Voltage:
    Power factor corrected input 0.98, 100-240Vac ±10%
    50/60Hz, 2A, maximum
  • X-Ray Tube Voltage:
    Nominal X-Ray tube voltage is adjustable between
    40kV to 100kV
  • X-Ray Tube Current:
    100uA to 1mA over specified tube voltage range
  • X-Ray Tube Power:
    100W, maximum continuous
  • Voltage Regulation:
  • Line:
    ±0.1% of maximum output voltage over a
    ±10% change of nominal input line voltage
  • Load:
    ±0.1% of maximum rated voltage for 100uA to 1mA load change
  • Voltage Accuracy:
    Voltage measured across the X-Ray tube is within ±2% of the programmed value
  • Voltage Risetime:
    Ramp time shall be 1 second from 10% to 90% of maximum rated output voltage
  • Voltage Overshoot:
    5% of maximum voltage, to return within 2.5% of maximum voltage in less than 50ms
  • Voltage Ripple:
    0.5% peak to peak of maximum voltage for frequencies

    1kHz
  • Emission Current Parameters Current Regulation:
  • Line:
    ±0.5% of rated output current over a
    ±10% change of nominal input line voltage
  • Load:
    ±0.5% of rated output current for a change from 50% to 100% of rated output voltage
  • Current Accuracy:
    Current measured through the X-Ray tube is within ±1% of the programmed value
  • Current Risetime:
    Ramp time shall be 1 second from 10% to 90% of maximum rated current
  • Arc Intervention:
    3 arcs in 10 seconds with a 200ms quench = Shutdown
  • Filament Configuration:
    Internal high frequency AC filament drive with closed loop filament emission control