Back
Browse Products By:
APPLICATIONS
Back
Browse Products By Application
View All Applications

PMX

  • Custom Designed Specifically for Mammography Applications
  • Compact Space Saving Modular Format
  • Fast Rise and Fall Times Help Minimize Unnecessary Patient Radiation Exposure
  • Dual Speed Starter, Boost/Brake Capability
  • RS-232 and Optional Ethernet Interfaces

Certifications & Compliance

Get a Quote

Spellman’s PMX is a high performance X-Ray generator designed specifically for mammography applications including FFDM and DBT. It features 2 and 3 point expo­sure modes as well as smart AEC with a pre-exposure.

The PMX is a 5kW high frequency X-Ray generator integrated with a dual filament power supply, and a dual-speed X-Ray tube starter. Due to the PMX's superior power conversion technology and inverter design it can provide stable and accurate X-Ray tube high voltage with fast rise and fall times.

Smooth system integration of the PMX is accomplished through the use of RS-232 and optional Ethernet digital in­terface, pre-loaded X-Ray tube parameters, easy access interlocks and I/O connections, and internal generator di­agnostics. Optional GUI utility software is also offered to help with initial system integration.

  • Specifications

Specifications

(Ref. 128113-001 REV. C)

  • Input Voltage:
    200-240Vac (±10%), single phase, 50Hz/60Hz
  • Input Current:
    Minimum 35A service recommended for 5kW operation
    External EMC Filter (Schaffner FN2070-36-08-36A) required to meet CE/EMC specifications – Not provided
    Mains Contactor – Not provided
    Customer is responsible for mains safety disconnection.
    Output (Tube) Voltage
  • Output Voltage Range:
    20kV to 49kV
  • Polarity:
    Positive, grounded cathode X-Ray tube
  • Accuracy:
    2% (measured per IEC60601-2-45)
  • Reproductibility:
    <0.5%
  • Rise Time:
    <1ms to within 98% of the programmed voltage
  • Fall Time:
    <10ms with a max HV cable length of 8 feet (2.4 meters)
  • Ripple:

    4% p-p
  • Output (Tube) Current/Power:

    Output Current Range:

    10mA to 200mA

    Output Power:

    5kW @ 0.1 second loading time,
    300 Watts maximum average power
  • Accuracy:
    < ±10% on exposure less than 10ms (measured per IEC60601-2-45)
  • Rise Time:
    <1ms to within 95% of the programmed mA value
  • Fall Time:
    <10ms with a maximum HV cable length of 8 feet (2.4 meters)
  • Exposure Time (Loading Time):

    Maximum Single Exposure Time:

    10 seconds

    Shortest Single Exposure Time:

    5ms
  • Loading time accuracy:
    ±3% +1ms (measured per IEC60601-2-45)
  • Maximum mAs:
    600mAs
  • Exposure Timer:
    5mS-10 seconds
  • Accuracy:
    < ±10% (measured per IEC60601-2-45)
  • Reproductibility:
    <0.5%
  • Filament Configuration:

    DC filament drive:

    self corrected filament preheat settings with closed loop emission control and smart learning algorithm

    Filament Output:

    0-6 amps at a compliance of 5.5 volts, maximum.
  • Dual Speed Starter:
    High speed (180Hz) and low speed (60Hz) can be
    configured via the serial interface.
    Boost and Brake capability provided.
  • High Voltage Connector:
    60kV, Claymount CA-3 type or equivalent
  • Communication Interface:
    RS-232 standard, optional Ethernet
  • Grounding Point:
    M5 ground stud provided on chassis
  • Environmental:
    Temperature Range:
    Operating: 10°C to 40°C
    Storage: -40°C to 85°C
    Humidity: 20% to 85%
    RH, non-condensing.
  • Cooling:
    Internal fan
  • Dimensions:
    9.47. H X 7.19. W X 13.72. D (240.5mm x 182.6mm x 348.5mm)
  • Weight:
    <23 pounds (10.5kg)
  • Regulatory Approvals:

    Designed to meet EMC:

    IEC 60601-1-2. UL/CUL recognized file E242584. RoHS compliant.
  • Application Features:
    2 point/3 point exposure modes
    AEC/Smart AEC exposure modes
    Dual Speed Starter
    X-Ray tube anode heat calculator
    Preloaded X-Ray tube parameters and expandable X-Ray tube library
Service Solution Brochure

Spellman Repair Centers, Extended Warranty Programs, Preventative Maintenance Programs, Calibration Services and more.

Download
2026 High Voltage Reference Manual

We compiled all of our useful FAQ’s, application notes, articles, technical papers and three glossaries into a single handy reference guide.

Download