XRB160PN480/2 Monoblock®
- Integrated HV Supply, Filament Supply, X-Ray Tube, Beam Port and Control Electronics
- Compact & LIghtweight
- Universal Input, Power Factor Corrected
- Can be Mounted in Any Physical Orientation
- Analog Control Interface and Standard RS-232 Digital Interface
160kV, 480W X-Ray Source
Spellman’s XRB160PN480/2 Monoblock® X-Ray source is designed for OEM applications powering its internal X-Ray tube up to 160kV at 480W. Features like universal input, small package size and a standard analog and RS-232 digital interface simplify integrating this Monoblock® into your X-Ray system. Standard models are available with fan shaped beam geometry. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability performance.
Typical applications:
- X-Ray Scanning: Food Inspection
- Fill Level Confirmation
- Security Applications
- Specifications
- Downloads
Specifications
(Ref. 128087-001 REV. J)
- X-Ray Characteristics:
- Tube Type:
Glass tube, Tungsten target, Be filter - Focal Spot:
0.8mm x 0.8mm (IEC336) - Beam Filter:
1mm curved window - Beam Geometry:
Asymmetrical fan 80° x 10° ±2° - Input Voltage:
100-240Vac ±10%, 50/60Hz, 6.5A max; 12Vdc @ 5A - X-Ray Tube Voltage:
Nominal X-Ray tube voltage is adjustable between
80kV to 160kV - X-Ray Tube Current:
0.25mA to 3mA over specified tube voltage range - X-Ray Tube Power:
500W, maximum - Voltage Regulation:
- Line:
±0.1% for a ±10% input line change of nominal input line voltage - Load:
±0.1% for a 0.3mA to 3mA load change - Voltage Accuracy:
Voltage measured across the X-Ray tube is within ±1% of the programmed value - Voltage Risetime:
Ramp time shall be <1 second from 10% to 90% of rated output - Voltage Overshoot:
Within 5% of rated voltage in <10ms - Voltage Ripple:
0.1% p p of rated voltage @ ≤1kHz - Current Regulation:
- Line:
±0.1% for a ±10% input line change of nominal input line voltage - Load:
±0.5% @ 80-160kV, 0.3mA to 3mA - Current Accuracy:
Current measured through the X-Ray tube is within ±1% of the programmed value - Current Risetime:
<1 second from 10% to 90% of rated output - Arc Intervention:
4 arcs in 10 seconds with a 200ms quench = Shutdown - Filament Configuration:
Internal high frequency AC filament drive with closed loop filament emission control