CT - Premium Series

* Ask about our subsystem capabilities

  • 110kW 128+ slice with <0.25s rotation to handle even the most demanding clinical applications
  • Sub ms rise and fall times to support Spectral Imaging and Dual Energy
  • Integrated multi-function grids for deflection, focus and cut off to control dose and provide the best image quality
  • Bi-polar or Anode End Grounded solutions
  • Compatible with all industry standard X-Ray tubes including Flat Emitter technology
Get a Quote

Premium X-Ray Generators for Rotating CT Scanner Applications

Spellman High Voltage has been designing, manufacturing and providing CT generators for OEM customers for over three decades. We have unrivaled experience being the first provider of a CT generator used in a commercially available continuous rotating CT scanner in 1985. Since then Spellman has enhanced performance and developed new CT generators year after year, setting the standard for CT generator technology.

Spellman High Voltage has manufactured over 30,000 CT generators. We have the knowledge, capability and experience to provide you with the CT generator technology you need.

  • Specifications
  • Models
  • Downloads

Specifications

  • Model Designation:
    CT140N110
  • Input Power:
    400Vac, 3Ø, +15%/-20%, 50/60Hz
  • Auxiliary Input Power:
    Not Required
  • kV Output Range:
    60kV - 140kV
    Anode grounded
  • kV Output Accuracy:
    ±0.5% of setting
  • mA Output Range:
    6mA to 900mA, 110kW maximum
  • mA Regulation/Accuracy:
    2% or 1mA, whichever is larger, for 20mA - 900mA
  • Filament Output:
    1 output: 10Vac @ 0-8A
  • Filament Accuracy:
    ±2% from 25% to 75%, 5% above
  • Filament Repeatability:
    ±0.2%
  • Starter:
    HSS configured to match X-Ray tube used
  • Interface:
    Custom serial interface
  • Rotation:
    Up to 240 rpm
  • Configuration:
    Four chassis
  • Grid Option:
    n/a
  • Magnetic Z Deflection Option:
    n/a

Models

ModelCT - ValueCT - PerformanceCT - Premium
 32/42/50kW50/80kW 110 kW
Model DesignationCCT70PN32 CCT70PN42 CCT70PN50CT70PN50 CT70PN80CT140N110
Input Power400Vac, 3Ø, +15%/-20%, 50/60Hz
Auxiliary Input Power220/230Vac, 1Ø
+15%/-10%, 50/60Hz
220/230Vac, 1Ø
+15%/-10%, 50/60Hz
Not Required
kV Output Range60kV - 140kV
(bipolar, 30kV – 70kV)
60kV - 140kV
(bipolar, 30kV – 70kV)
60kV - 140kV
Anode grounded
kV Output Accuracy±0.5% of setting
mA Output Range10mA – 350mA,
42kW maximum
10mA – 420mA / 50kW maximum
10mA – 667mA / 80kW maximum
6mA to 900mA,
110kW maximum
mA Regulation/Accuracy

2% or 2mA or which is
er for 20mA -350mA

2% or 1mA or which is
larger for 20mA -420mA
2% or 1mA or which is
larger for 20mA -900mA
Filament Output2 outputs: 15Vac @ 0-6A
up to 8A available
1 output: 15Vac @ 0-6A,
up to 15A available
1 output: 10Vac @ 0-8A,
Filament Accuracy±5% of setting±5% of setting±2% from 25% to 75%, 5% above
Filament Repeatability±0.2%
StarterHSS configured to match
X-Ray tube used
InterfaceCustom serial interface
RotationUp to 120 rpmUp to 120 rpmUp to 240 rpm
ConfigurationSingle ChassisTwo chassisFour chassis
Grid Optionn/aDual grid/single filament, 0-4kV @ 0-4kHzn/a
Magnetic Z