뒤로
다음 기준으로 제품 찾아보기:
응用 분야
뒤로
응용 분야별 제품 보기
모든 응용 분야 보기

XRB80N100 Monoblock®

  • Integrated HV Supply, Filament Supply, X-Ray Tube, Beam Port and Control Electronics
  • Compact & Lightweight
  • Universal Input, Power Factor Corrected with Internal EMI Filter
  • Can be Mounted in Any Physical Orientation
  • Analog Control Interface and Standard RS-232 Digital Interface

Certifications & Compliance

견적 받기

80kV, 100W X-Ray Source

Spellman’s XRB80N100 Monoblock® X-Ray source is designed for OEM applications powering its internal X-Ray tube up to 80kV at 100W. Features like universal input, small package size and a standard analog and RS-232 digital interface simplify integrating this unit into your X-Ray system. The XRB80N100 is available either with fan shaped (standard) or (optional) cone shaped beam geometries. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability performance.

Typical XRB80N100 Monoblock® Integrated X-Ray Source applications:

  • X-Ray Scanning: Bone Densitometry
  • Thickness Measurement
  • Food Inspection
  • Fill Level Confirmation
  • Parcel Inspection
  • 사양
  • 다운로드

사양

(Ref. 128065-001 REV. Y)

  • X-Ray Characteristics:
  • Focal Spot:
    0.5mm (IEC 336)
  • Beam Filter:
  • Ultem:
    3.30mm ±0.15mm
  • Oil:
    8mm ±0.1mm
  • Glass:
    1.8mm ±0.25mm
  • BE:
    0.8mm
  • Beam Geometry:
  • Fan:
    Standard. The beam angular coverage will be 75° with the beam plane perpendicular to the X-Ray tube axis and 13° wide.
  • Cone:
    Optional. 25° cone beam
  • Input Voltage:
    Power factor corrected input 0.98, 100-240Vac ±10%
    50/60Hz, 2A, maximum
  • X-Ray Tube Voltage:
    Nominal X-Ray tube voltage is adjustable up to 80kV
  • X-Ray Tube Current:
    150uA to 1.25mA over specified tube voltage range
  • X-Ray Tube Power:
    100W, maximum continuous
  • Voltage Regulation:
  • Line:
    ±0.05% of maximum output voltage over a ±10% change of nominal input line voltage
  • Load:
    ±0.1% of maximum rated voltage for 150uA to 1.25mA load change
  • Voltage Accuracy:
    Voltage measured across the X-Ray tube is within ±2% of the programmed value
  • Voltage Risetime:
  • Standard:
    Ramp time shall be 500ms from 10% to
    90% of maximum rated output voltage
  • Optional:
    5 seconds. Specify at time of order
  • Voltage Overshoot:
    5% of maximum voltage, to return within 2.5% of
    maximum voltage in less than 100ms
  • Voltage Ripple:
    1% peak to peak of maximum voltage for
    frequencies

    1kHz
    Emission Current Parameters
  • Current Regulation:
  • Line:
    ±0.05% of rated output current over a ±10% change of nominal input line voltage
  • Load:
    ±0.1% of rated output current for a change from 50% to 100% of rated output voltage
  • Current Accuracy:
    Current measured through the X-Ray tube is within ±2% of the programmed value
  • Current Risetime:
  • Standard:
    Ramp time shall be 500ms from 10% to 90% of maximum rated current
  • Optional:
    5 seconds. Specify at time of order
  • Arc Intervention:
    3 arcs in 10 seconds with a 200ms quench = Shutdown
  • Filament Configuration:
    Internal high frequency AC filament drive with closed loop filament emission control