뒤로
다음 기준으로 제품 찾아보기:
응用 분야
뒤로
응용 분야별 제품 보기
모든 응용 분야 보기

XRB160PN688/CT Monoblock®

  • Integrated HV Supply, Filament Supply, X-Ray Tube, Beam Port and Control Electronics
  • Compact & LIghtweight
  • Power Factor Corrected
  • Can be Mounted in Any Physical Orientation
  • Standard RS-232 Digital Interface

*Note: All specifications are subject to change without notice. Please consult the English PDF version of this datasheet for the most up-to-date revision.

견적 받기

160kV, 688W X-Ray Source

Spellman’s XRB160PN688/CT Monoblock® X-Ray source is designed for OEM applications powering its internal X-Ray tube up to 160kV at 688W. Features like small package size and RS-232 digital interface simplify integrating this unit into your X-Ray system. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability performance.

Typical applications:

  • X-Ray Scanning: Food Inspection
  • Fill Level Confirmation
  • Security Applications
  • 사양
  • 다운로드

사양

(Ref. 128088-001 REV. G)

  • X-Ray Characteristics:
  • Tube Type:
    Glass tube, Tungsten target, Be filter
  • Focal Spot:
    2.5mm x 2.5mm (IEC 60336)
  • Beam Filter:
    1.5mm glass, 9-26mm oil, 0.4mm Al
  • Beam Geometry:
    Asymmetrical fan 93° x 15°
  • Input Voltage:
    230Vac, ±15%, 50/60Hz, 5A max
  • X-Ray Tube Voltage:
    Nominal X-Ray tube voltage is adjustable between 50kV to 160kV
  • X-Ray Tube Current:
    0.25mA to 4.3mA over specified tube voltage range
  • X-Ray Tube Power:
    700W, maximum continuous
  • Voltage Regulation:
  • Line:
    ±0.1% ±10%
  • Load:
    ±0.1% 0.25 to 6mA
  • Voltage Accuracy:
    Voltage measured across the X-Ray tube is within 1% of the programmed value
  • Voltage Risetime:
    Ramp time shall be 500ms nominal, 1 second maximum from 10% to 90% of rated output
  • Voltage Overshoot:
    Within 5% of rated voltage
  • Voltage Ripple:
    1% p-p of rated voltage
  • Current Regulation:
  • Line:
    ±0.5% ±10%
  • Load:
    ±0.5% 0.25 to 4.3mA
  • Current Accuracy:
    Current measured through the X-Ray tube is within 5% of the programmed value
  • Current Risetime:
    500ms nominal, 1 second maximum
  • Arc Intervention:
    4 arcs in 10 seconds = shutdown
  • Filament Configuration:
    Internal high frequency AC filament drive with closed loop filament emission control
  • Digital Interface:
    RS-232 interface.
  • Control Software:
    A demo GUI for engineering evaluations will be provided for the RS-232 digital interface upon request.
  • Interlock Signals:
    A hardware interlock function is provided
  • Operating Temperature:
    5°C to +40°C
  • Storage Temperature:
    -25°C to +65°C
  • Humidity:
    10% to 90% relative humidity, non-condensing
  • Cooling:
    Heat exchanger w/fan and oil pump, powered from AC
  • Input Line Connector:
    5 pin Molex 26-60-4050
  • Cooling Power Connector:
    4 pin Tyco 206061-1
  • Digital Interface Connector:
    9 pin D connector, female
  • Grounding Point:
    M5 ground stud provided on chassis
  • Dimensions:
    See drawing
  • Weight:
    140lbs ±10lbs (64kg ±4.5kg)
  • Orientation:
    Can be mounted in any orientation.
  • X-Ray Leakage:
    Less than 300uR/hr at a distance of 10cm from all surfaces
  • Special Features:
    Stationary or rotating CT application up to 100rpm at a maximum radius 20.25˝ (514mm).