Spellman’s MMB125PN3.5 Medical Monoblock® with <1ms rise time is ideal for extremities, specialized applications and vascular imaging. Benefit from Spellman’s 70 plus years of high voltage innovation with this MMB125PN3.5 Monoblock® X-Ray source specifically designed for the C-Arm OEM. Our advanced technology and design expertise in conjunction with our depth of manufacturing capabilities position Spellman as the wise choice for your Monoblock® requirements.

Spellman’s Ultra-Fast MMB Series utilizes our unique technologies that increases the quality of the X-Ray beam and significantly reduces patient dose. Pulsed Fluoroscopy enhances imaging of dense and complex anatomy to further support dose management. With a rise time of <1ms, Spellman’s MMB Series is among the fastest in the industry.

  • 사양

사양

(Ref. 128143-001 REV. A)

  • X-Ray Characteristics:

    Focal Spot:

    0.5mm for small focus
    1.6mm for large focus

    Target Angle: 
    16°
    Target Material: 
    Tungsten
    Beam Filter: 
    0.8mm Al
    X-Ray Leakage: 
    Less than <1mGy/hour @ 1meter from tank surface.

    Maximum Filament Current:
    4.3A
    Filament Inverter

    Switching Frequency:

    20kHz

    Anode Heat Content:

    35.5kJ
    Maximum Anode

    Heat Dissipation:

    600W
    Distance Between
    Focal Spot to X-Ray

    Output Window:

    45mm
  • Input Power:

    Input Voltage:

    190-264Vac, single phase, 50/60 Hz

    Continuous Current:

    ≤9A @ 600W, 220V

    Peak Current:

    ≤34A @ 3500W, 220V
  • X-Ray Tube Voltage:

    Operational Range:

    40 to 125kV

    kV Rise Time:

    <1ms (from 10% to 90%)

    Reproducibility:

    ≤0.05 per IEC60601-2-54

    kV Ripple:

    ≤1%
    kV Accuracy

    Fluoroscopy:

    ±5%

    kV Overshoot:

    ≤5%
  • X-Ray Tube Current:

    Fluoro:

    0.2mA to 6 mA (small focal spot)

    Pulsed Fluoro:

    0.5mA to 10 mA (small focal spot)

    Radiography:

    10mA to 40mA (large focal spot)

    mA Accuracy:

    ±10%

    Maximum Operating Conditions:

    Continuous

    Low Level Fluoro:

    6 mA for 5 minutes
    Continuous

    High Level Fluoro:

    10 mA for 2 minutes

    Pulsed Fluoro:

    0.5fps, 1fps, 2fps, 4fps, 8fps, 15fps,
    25fps, 30fps for 2 minutes

    Pulse on time:

    the lower value of 40ms
    or 50% duty cycle

    Radiographic:

    Single shot large focal spot,
    0.1 seconds @ 3500 watts
    Maximum Anode

    Heat Dissipation:

    600W

    Average Power:

    600W for 5 minutes

    Tank Heat Content:

    No less than 675kHU

    Max. Cooling Rate:

    150W
  • X-Ray Tube Current Protection:

    Over Current (High mA):

    Trip point is set at 15% over maximum rated current, or at 15% over programmed output current value for longer than 50ms.

    Under Current (High mA):

    Trip point is set at 15% under programmed output current value for longer than 50ms.

    Over Current (Low mA):

    Trip point is set at 20% over maximum rated current, or at 20% over programmed output current value for longer than 50ms.

    Under Current (Low mA):

    Trip point is set at 20% under programmed output current value for longer than 50ms.
  • X-Ray Tube Voltage Protection:

    Over Voltage (High kV):

    Trip point is when the kV output exceeds 137.5kV for more than 10ms or it set is at 10% over programmed output voltage value for longer than 30ms.

    Under Voltage (Low kV):

    Trip point is set at 10% under the programmed output voltage value for longer than 30ms.
  • X-Ray Tube Arc Protection:

    Arc Intervention:

    Unit will detect an Arc but HV will not be shut down, but if multiple arcs occur (4 arcs in 10 seconds) the unit will shut down.
  • Over Temperature Protection:

    Over Temperature:

    Over temperature of tank cover trip point shall be within 60°C ±5°C. Over temperature of tank oil trip point shall be set 65°C.
  • Operating Temperature:
    0 to +40 ºC.
  • Storage Temperature:
    -20 to +70 ºC
  • Altitude:
    0 to 8000 feet (0 to 2438 meters)
  • Humidity:
    5 to 95%, non-condensing.
  • Dimensions:

    X-Ray Tank:

    12.2” x 6.2” x 5.7” (310mm x 158mm x 145mm)

    Controller:

    14.9” x 10.6” x 3.2” (380mm x 270mm x 83mm)
  • Weight:

    X-Ray Tank:

    27.5 pounds (13kg)

    Controller:

    9.9 pounds (4.5kg)
  • Regulatory Approvals:

    Designed to be compliant to:

    IEC 60601-1:

    2005+A

    Medical electrical equipment Part 1:

    General requirements for basic safety and essential performance.
    IEC 60601-1-2

    2007 Medical electrical equipment Part 1-2:

    General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility—Requirements and tests.
    IEC 60601-1-3

    2008 Medical electrical equipment Part 1-3:

    General requirements for basic safety and essential performance- Collateral standard: Radiation protection in diagnostic X-Ray equipment.
    IEC 60601-2-54

    2009 Medical electrical equipment Part 2:

    Particular requirements for the basic safety and essential performance of X-Ray equipment for radiography and radioscopy.
    EMC

    Designed to meet IEC requirements for medical components.
    (Note:External EMI filter may be required) RoHS
    Controller and Tank Assembly are RoHS compliant.