  • Integrated HV Supply, Filament Supply, X-Ray Tube, Beam Port and Control Electronics
  • Compact & Lightweight
  • Can be Mounted in Any Physical Orientation
  • Analog or Digital Control Interface

Certifications & Compliance

견적 받기

80kV, 20-50W X-Ray Source

Spellman’s XRB011 Series of Monoblock® X-Ray sources are designed for OEM applications powering its internal X-Ray tube up to 80kV at 20W and 80kV at 50W. Features like 24Vdc input voltage, small package size, standard analog interface and RS-232/Ethernet digital interface simplify integrating the XRB011 into your X-Ray system. Proprietary emission control circuitry provides excellent regulation of X-Ray tube current, along with outstanding stability and performance.

Typical applications:

  • Medical X-Ray: Fluoroscopy and Radiography for Extremities, Specimen Radiography.
  • Pulsed Fluoroscopy (contact Spellman sales)
  • Industrial X-Ray: Component inspection and Non-Destructive Testing
  • 사양
  • 다운로드

사양

(Ref. 128090-001 REV. AF)

  • X-Ray Characteristics:
  • Tube Type:
    Micro focus tube
  • Focal Spot:
    33µm Nominal, 50µm max. (IEC 336)
  • Beam Filter:
    Ultem 0.060. (1.5mm)
    Oil 0.175. (4.4mm)
  • Beam Geometry:
    Symmetrical 40° cone
  • Input Voltage:
  • 20W:
    24Vdc ±1V @ 2.5A
  • 50W:
    24Vdc ±1V @ 4A
  • X-Ray Tube Voltage:
    Nominal X-Ray tube voltage is adjustable between 35kV to 80kV
  • X-Ray Tube Current:
  • 20W:
    0-250µA over specified tube voltage range
  • 50W:
    0-700µA over specified tube voltage range
  • X-Ray Tube Power:
    20/50W, maximum continuous
  • Voltage Regulation:
  • Line:
    ±0.5% for a ±1V change of nominal input line voltage
  • Load:
    ±0.1% for a load change of 25µA to maximum rated current
  • Voltage Accuracy:
    Voltage measured across the X-Ray tube is within ±1% of the programmed value
  • Voltage Risetime:
    Ramp time shall be =250ms from 10% to 90% of maximum rated output voltage
  • Voltage Temperature Coefficient:
    ≤100ppm/°C
  • Over Temperature Fault:
    Indicates that the internal oil temperature has exceeded 65° C. The high voltage output will be disabled. Toggling the X-Ray ON Command OFF and ON will reset the fault.
  • Over Voltage Fault:
    An overvoltage (OV) fault is detected when the output voltage exceeds 82kV. The high voltage output will be disabled. Toggling the X-Ray ON Command OFF and ON will reset the fault.
  • Voltage Ripple:
    1% peak to peak
  • Current Regulation:
  • Line:
    ±0.5% for a ±1V change of nominal input line voltage
  • Load:
    ±0.5% for a voltage change of 35kV to 80kV
  • Current Accuracy:
    Current measured through the X-Ray tube is within ±2.5% of the programmed value
  • Over Current Fault:
    An overcurrent (OC) fault is detected when the emission current exceeds 275µA (20W model) and 710µA (50W model). Toggling the X-Ray ON Command OFF and ON will reset the fault.
  • Arc Intervention:
    One arc fault. The high voltage output will be disabled. Toggling the X-Ray ON command OFF and ON will reset the fault.
  • Filament Configuration:
    Internal high frequency AC filament drive with closed loop filament emission control