뒤로
다음 기준으로 제품 찾아보기:
응用 분야
뒤로
응용 분야별 제품 보기
모든 응용 분야 보기

DGM945

  • Multiple Outputs
  • Three Control Inputs
  • Three Optical Encoders
  • RS-232 Serial Port for Control and Monitoring
  • OEM Customization Available

Certifications & Compliance

견적 받기

Modular High Voltage Supply for Image Intensifiers

Spellman’s new DGM945 high voltage power supply for Image Intensifier applications continues to set the stan­dards for high voltage power conversion technology. The DGM945 also has inputs to allow measurement ofthe photocathode current and getter current. The photo-cathode measurement has two ranges : sensitive range: 0 to 50nA and a standard range: 0 to 15µA. The sensitiveinput is used to provide a monitor output with a sensitivityof 0.1V/nA. The getter input allows measurement of the getter current, indicating if the Image Intensifier tube is still holding a good vacuum and is without gases causing ions.

Monitoring of all output voltages is possible via local test points or over the serial interface. The unit can also storeup to eight modes of operation. The output voltages are independently adjustable in each mode either by the rotary encoders built into the unit or via the serial interface.

The DGM series can be adapted to suit specific requirements with a wide selection of multiple output voltages and power capabilities in a compact package, making it perfect for the OEM user.

  • 사양

사양

(Ref. 128114-001 REV. C)

  • Input Voltage:
    +24Vdc ±10%
  • Input Current:
    500mA maximum
  • Programmable Output Voltages:

    Anode:

    Output Voltage: 30kV
    Output Current: 5µA continuous (50µA peak)

    Electrode 1:

    Output Voltage: 10V to 350V
    Output Current: 0 to 10µA

    Electrode 2:

    Output Voltage: 200V to 2kV
    Output Current: 0 to 10µA

    Electrode 3:

    Output Voltage: 2kV to 20kV
    Output Current: 0 to 20µA

    Electrode 4:

    Output Voltage: 2.7kV
    Output Current: 0 to 10µA
  • Ripple:
    <0.3%
  • Temperature Coefficient:
    <200ppmK-1
  • Stability:
    <0.3% over 8 hrs
  • Temperature:
    Operating: +5°C to +55°C
  • Input Connector:
    15 pin male sub D
  • Dimensions:
    5.47”H x 5.67”W x 1.97”D (139mm x 144mm x 50mm)
  • Weight:
    2.86 lb. (1.3kg)
  • Regulatory Approvals:
    Compliant to EEC EMC Directive. Compliant to EEC Low Voltage Directive.

Typical applications

  • Radiology
  • Cardiology
  • Neuroradiology
  • Night Surveillance
  • Non Destructive X-Ray Inspection
  • Image Intensifiers
서비스 솔루션 책자

Spellman 수리 센터, 연장 보증 프로그램, 예방 유지 보수 프로그램, 교정 서비스 등을 제공합니다.

다운로드
2026 고전압 참조 매뉴얼

우리는 유용한 FAQ, 응용 노트, 기사, 기술 논문 및 용어집 3 개를 모두 하나의 편리한 참조 안내서로 만들었습니다.

다운로드