SpellmanHV CZE2000 Auto-Reversing High Voltage Power Supply

CZE2000

  • Cost Effective Modular Supply; Ideal for Electrospinning
  • 0-30kV. 0-300µA; Local or Remote Programmable
  • Polarity Reversible Upon Command in <1 Sec at No Load
  • Low Stored Energy, Current Limited Output

*Note: All specifications are subject to change without notice. Please consult the English PDF version of this datasheet for the most up-to-date revision.

RoHs Compliant logo CE RU Marking

CZE2000 Auto-Reversing Modular High Voltage Power Supply

Spellman’s CZE2000 modular high voltage power supply is ideal for OEM usage. It is specifically designed to meet the needs of applications requiring a hot switched reversible output voltage. The output polarity of the unit can be quickly and safely reversed via the Polarity Control Signal provided on the interface connecter.

Both the output voltage and current are fully adjustable via ground referenced remote programming signals such that 0 to 10Vdc corresponds to 0 to 100% rated output voltage and current.

Remote motioning functionality is provided by voltage and current test points such that 0 to 10Vdc corresponds to 0 to 100% rated voltage and current. Additionally remote polarity and mode indicators provide a comprehensive overview of power supply operation.

Excellent load and line regulation specifications along with outstanding stability and low ripple assure a stable high voltage output for consistent process results.

Typical applications:

  • Electrospinning
  • Mass Spectrometry
  • Capillary Electrophoresis
  • Electrostatic Research

사양

(Ref. 128076-001 REV. L)

Input Voltage:

24Vdc, ±10%

Input Current:

Less than 1 amp

Efficiency:

75% typical

Output Voltage:

See selection table

Output Current:

See selection table

Polarity:

Auto reversible via command

Power:

10 watts, maximum

Line Regulation:

0.01% for a 10% input voltage change

Load Regulation:

0.01% for a full load change

Ripple:

0.1% Vp-p

Stability:

0.02% per 8 hours (after 1/2 hr warmup)

NL Time Constant:

100ms

Stored Energy:

0.2 Joules at 30kV

Temperature Coefficient:

100ppm/°C

Operating Temperature:

0°C to 40°C

Storage Temperature:

-40°C to 85°C

Humidity:

10% to 85% RH, non condensing

Cooling:

Convection cooled

Dimensions:

3.5”H x 5”W x 10”D (8.9cm x 12.7cm x 25.4cm).

Weight:

6.2lbs. (2.8kg)

Interface Connector:

25 pin male D connector

HV Output Connector:

Detachable 36. (0.91m) cable provided

Regulatory Approvals:

Compliant to EEC EMC Directive. Compliant to EEC Low Voltage Directive. UL/CUL recognized file E148969. RoHS Compliant.

CZE2000 SELECTION TABLE

Maximum Rating Model Number
kV mA  
5 2.0 CZE5PN2000
10 1.0 CZE10PN2000
15 0.67 CZE15PN2000
20 0.50 CZE20PN2000
30 0.30 CZE2000

 

CZE2000 25 PIN MALE D CONNECTOR

Pin Signal Parameters
1 +24Vdc Return Power Return
2 +24Vdc Return Power Return
3 +24Vdc Return Power Return
4 HV Enable/Inhibit Open or <1Vdc = HV OFF, >3.4Vdc (up to 15Vdc) = HV ON1
5 Voltage Test Point 0 to 10Vdc = 0 to 100% rated output, Zout =10kΩ, 1%
6 Current Test Point 0 to 10Vdc = 0 to 100% rated output, Zout =10kΩ, 1%
7 Chassis Ground Ground
8 Remote Voltage Control 0 to 10Vdc = 0 to 100% Rated Output, Zin =10MΩ
9 Remote Current Control 0 to 10Vdc = 0 to 100% Rated Output, Zin =10MΩ
10 +10Vdc Reference Output +10Vdc, 4mA maximum
11 Signal Return Signal Return
12 Polarity Control Open or >3.4Vdc (up to 15Vdc) = Positive Polarity. Grounded or <1Vdc = Negative Polarity
13 Positive Polarity Indicator +24Vdc sourced through a 100Ω series limiting resistor. +24Vdc = active signal
14 +24Vdc Input Power Input
15 +24Vdc Input Power Input
16 Chassis Ground Ground
17 Negative Polarity Indicator +24Vdc sourced through a 100Ω series limiting resistor. +24Vdc = active signal
18 I Mode Indicator Open collector pulled up internally to +15Vdc through 2.7kΩ resistor with a 470Ω limiting resistor in series. Transistor OFF = signal active
19 V Mode Indicator Open collector pulled up internally to +15Vdc through 2.7kΩ resistor with a 470Ω limiting resistor in series. Transistor OFF = signal active
20 Return Current Test Point 0 to 10Vdc = 0 to 100% rated output current, as measured returned from load. Zout =10kΩ, 1%
21 Load Return High Voltage Return Point. Required for GFI circuit functionality
22 Ground Fault Indicator Open collector pulled up internally to +15Vdc through 4.7kΩ resistor with a 470Ω limiting resistor in series. Transistor OFF = signal active
23 Spare No Connection